Yokogawa EJX Differential Pressure Transmitter
Type | Data source |
---|---|
EJX110A | Product Data Sheet GS 01C25B01-01E November 2019 |
EJX130A | Product Data Sheet GS 01C25B04-01EN May 2014 |
EJX910A | Product Data Sheet GS 01C25R01-01EN April 2012 |
EJX930A | Product Data Sheet GS 01C25R04-01EN February 2014 |
Reference Accuracy U(ac)
All Models
for Spans below…
Capsule | X |
---|---|
L | 2 kPa |
M | 10 kPa |
H | 70 kPa |
U(ac) = ± 0.04% of Span
Otherwise,
Capsule | |
---|---|
L | |
M | |
H |
(for EJX130 & EJX930: capsule M & H only)
Drift (stability) U(d)
All Models
± 0.1% of URL for 10 years
Power supply effect U(ps)
All Models
± 0.005% of calibrated span per volt change.
Ambient temperature effect U(ac)
EJX110A & EJX910A
Capsule | |
L | ±(0.09% URL + 0.055% Span) |
M | ±(0.009% URL + 0.04% Span) |
H | ±(0.0125% URL + 0.04% Span) |
EJX130A & EJX930A
Capsule | |
M | ±(0.009% URL + 0.07% Span) |
H | ±(0.0125% URL + 0.07% Span) |