PMP75 Endress & Hauser Pressure Transmitter
Type | Data source |
---|---|
PMP75 Standard | Product Data Sheet 2.5M309CW 2012 |
PMP75 Platinum | Product Data Sheet 2.5M309CW 2012 |
Reference Accuracy U(ac)
Standard
± 0.2% of calibrated span
Platinum
± 0.2% of calibrated span
Drift (stability) U(d) =
Standard
± 0.1% of URL for 1 year
Platinum
± 0.1% of URL for 1 year
Power supply effect U(ps)
± 0.005% of calibrated span per volt change.
Ambient temperature effect U(ac)
Standard
Range | -10°C to 60°C | -40°C to -10°C 60°C to 85°C |
---|---|---|
400 mbar |
|
|
1, 2, 4, 10 & 40 bar |
|
|
100 bar |
|
|
400 bar |
|
|
700 bar |
|
|
Platinum
Range | -10°C to 60°C | -40°C to -10°C 60°C to 85°C |
---|---|---|
400 mbar |
|
|
1, 2, 4, 10 & 40 bar |
|
|