Type | Data source |
---|---|
PMP71 Standard | Product Data Sheet 2.5M309CW 2012 |
PMP71 Platinum | Product Data Sheet 2.5M309CW 2012 |
Reference Accuracy U(ac)
Standard
Range | Sensor | Low Turndown | High Turndown |
---|---|---|---|
400 mbar | Gauge & Absolute | 1:1 ± 0.05% |
> 1:1 |
1 bar | Gauge & Absolute | 1:1 to 2.5:1 ± 0.05% |
> 2.5:1 |
2 bar | Gauge & Absolute | 1:1 to 5:1 ± 0.05% |
> 5:1 |
4 bar | Gauge & Absolute | 1:1 to 10:1 ± 0.05% |
> 10:1 |
40 bar | Gauge & Absolute | 1:1 to 15:1 ± 0.075% |
> 15:1 |
100 bar | Gauge & Absolute | 1:1 to 10:1 ± 0.075% |
> 10:1 |
400 bar 700 bar |
Gauge & Absolute Absolute |
1:1 to 5:1 ± 0.15% |
> 5:1 |
Platinum
Range | Sensor | U(ac) |
---|---|---|
400 mbar & 1 bar 2, 4 & 10 bar |
Gauge & Absolute Absolute |
± 0.035% |
2, 4 & 10 bar | Gauge | ± 0.025% |
40 bar 100 bar |
Gauge & Absolute Absolute |
± 0.05% |
400 & 700 bar | Gauge & Absolute | ± 0.075% |
Drift (stability) U(d)
for 1 year | for 5 years | for 10 years | |
---|---|---|---|
Range | % of URL | ||
1 bar | ± 0.02% | ± 0.08% | ± 0.18% |
10 bar | ± 0.025% | ± 0.05% | ± 0.075% |
40 bar | ± 0.025% | ± 0.075% | ± 0.1% |
100 bar | ± 0.05% | ± 0.15% | ± 0.2% |
400 bar | ± 0.05% | ± 1.0% |
Power supply effect U(ps)
± 0.0006% of calibrated span per volt change.
Ambient temperature effect U(ac)
Standard
Range | -10°C to 60°C | -40°C to -10°C 60°C to 85°C |
---|---|---|
400 mbar | ||
1, 2, 4, 10 & 40 bar | ||
100 bar | ||
400 bar | ||
700 bar |
Platinum
The % of the set span for platinum is 5x the Standard values