Honeywell ST3000 Differential Pressure Transmitter
Type | Data source |
---|---|
100 | Product Data Sheet 34-ST-03-62 DP January 2006 |
900 | Product Data Sheet 34-ST-03-67 DP January 2006 |
Reference Accuracy U(ac)
100
Digital
Range | Low Turndown | High Turndown |
---|---|---|
120 & 130 | ± 0.0625% of calibrated span | |
120 (HA) | ± 0.035% of calibrated span | |
125 | ± 0.05% of calibrated span |
Analogue
Range | Low Turndown | High Turndown |
---|---|---|
120 & 130 | ± 0.075% of calibrated span | |
120 (HA) | ± 0.0375% of calibrated span | |
125 | ± 0.075% of calibrated span |
900
Digital
Range | Low Turndown | High Turndown |
---|---|---|
924 & 930 | ± 0.0625% of calibrated span |
Analogue
Range | Low Turndown | High Turndown |
---|---|---|
924 & 930 | ± 0.075% of calibrated span |
Drift (stability) U(d)
100
Range | |
---|---|
120 & 120(HA) | ± 0.01% of URL for 1 year |
125 | ± 0.015% of URL for 1 year |
130 | ± 0.04% of URL for 1 year |
900
Range | |
---|---|
924 | ± 0.015% of URL for 1 year |
930 | ± 0.04% of URL for 1 year |
Power supply effect U(ps)
All Models
Less than ± 0.005% of calibrated span per volt change.
Ambient temperature effect U(ac)
100
Digital
Range | Low Turndown | High Turndown |
---|---|---|
120 & 125 | ± 0.075% of calibrated span | |
130 | ± 0.075% of calibrated span |
Analogue
Range | Low Turndown | High Turndown |
---|---|---|
All Ranges | ± 0.1% of calibrated span |
900
Digital
Range | Low Turndown | High Turndown |
---|---|---|
924 & 930 | ± 0.225% of calibrated span |
Analogue
Range | Low Turndown | High Turndown |
---|---|---|
924 & 930 | ± 0.25% of calibrated span |