Foxboro IGP10 Pressure Transmitter

Type Data source
IGP10 Analogue Product Data Sheet PSS 2A-1C13 C
IGP10 HART Product Data Sheet PSS 2A-1C13 B
IGP10 Fieldbus Product Data Sheet PSS 2A-1C13 E

Reference Accuracy U(ac)

Analogue

± 0.2% of calibrated span

HART

Range Spans < 10% URL Spans > 10% URL
E & F ± 0.08% of calibrated span
All other ranges ± 0.06% of calibrated span

Fieldbus

Range Spans > 10% URL
E & F ± 0.07% of calibrated span
All other ranges ± 0.05% of calibrated span

Drift (stability) U(d)

All Models

± 0.05% of URL for 5 years

Power supply effect U(ps)

All Models

± 0.005% of calibrated span per volt change.

Ambient temperature effect U(ac)

Analogue

±(0.08% URL + 0.1% Span)

HART

±(0.03% URL + 0.06% Span)

Fieldbus

±(0.08% URL + 0.025% Span)

 

 