C260 – ADC Input

Description

This calculation performs an ADC check which can be used for the calibration of equipment where there is a linear relationship between the input quantity and the output.

References

Kelton calculation reference C260

KIMS calculation reference K122

Options

Transmitter Input

  • Current
  • Voltage

This option allows user to select whether the input is a current or a voltage.

Transmitter Output

  • Scalar
  • Factor
  • Pressure
  • Temperature
  • BS&W
  • Volume flow rate
  • Mass flow rate

This option allows user to select the relevant output type.

Measure input as voltage

Allows user to input a voltage which is then converted to a current using Ohm’s law.

Measure input as current

Allows user to input a current which is then converted to a voltage using Ohm’s law.

Square root extraction

This option is selected if the input has a square root relationship with the output.

Linearise as input approaches LRV

This option is selected to linearise the square root extraction function for values very close to or below the lower range value.

Calculation

Output Value

The output value is calculated by:

Or when square root extraction is applied:

When the input is less than 0.6% the input range the above linear function is used. When the input is above 0.8% of the input range the normal square root extraction function, as detailed above, is used. When square root extraction is linearised the following function is used when the input is between 0.6 and 0.8% of the input range.

Where
IP = Transmitter input value
IPmax = Transmitter maximum input value
IPmin = Transmitter minimum input value
OPmax = Upper range output value (which corresponds to the set maximum input value) for the instrument being calibrated
OPmin = Lower range output value (which corresponds to the set minimum input value) for the instrument being calibrated

 

 