Autrol APT3100 Pressure Transmitter

Type Data source
3100 Product Data Sheet C3100-E01B

Reference Accuracy U(ac)

Range for 0.1URL < Span < URL for 0.01URL < Span < 0.1URL
3 to 7 ± 0.075% of span
All other ranges ± 0.1% of span

Drift (stability) U(d)

± 0.125% of URL for 1 year

Power supply effect U(ps)

± 0.005% of calibrated span per volt change.

Ambient temperature effect U(ac)

Span < 10% of URL Span > 10% of URL
±(0.019% URL + 0.125% Span) ±(0.025% URL + 0.125% Span)

 

 

Back