ABB 2600T/264 Pressure Transmitter
Type | Data source |
---|---|
264DS | Product Data Sheet SS/264XS_6 2006 |
264DF | Product Data Sheet SS/264XF_5 2006 |
264BS | Product Data Sheet SS/264BS_4 2006 |
Reference Accuracy U(ac)
264DS
Range | Low Turndown | High Turndown |
---|---|---|
A | For spans less than 4:1 ± 0.1% of calibrated span |
For spans from 4:1 to 20:1 |
B | For spans less than 10:1 ± 0.1% of calibrated span |
For spans from 10:1 to 20:1 |
All other ranges | For spans less than 15:1 ± 0.075% of calibrated span |
For spans from 15:1 to 60:1 |
264DF
Range | Low Turndown | High Turndown |
---|---|---|
B | For spans less than 10:1 ± 0.1% of calibrated span |
For spans from 10:1 to 20:1 |
E | For spans less than 15:1 ± 0.075% of calibrated span |
For spans from 15:1 to 30:1 |
All other ranges | For spans less than 15:1 ± 0.075% of calibrated span |
For spans from 15:1 to 60:1 |
264BS
Range | Low Turndown | High Turndown |
---|---|---|
C | For spans less than 15:1 ± 0.075% of calibrated span |
For spans from 15:1 to 30:1 |
All other ranges | For spans less than 15:1 ± 0.075% of calibrated span |
For spans from 15:1 to 60:1 |
Drift (stability) U(d)
All Models
± 0.15% of URL per 5 years
Power supply effect U(ps)
All Models
± 0.005% of calibrated span per volt change.
Ambient temperature effect U(ac)
264DS
Range | |
---|---|
A | ±(0.1% URL + 0.1% Span) |
B | ±(0.06% URL + 0.1% Span) |
All other ranges | ±(0.04% URL + 0.065% Span) |
264DF
Range | |
---|---|
B | ±(0.06% URL + 0.1% Span) |
All other ranges | ±(0.04% URL + 0.065% Span) |
264BS
Range | |
---|---|
C | ±(0.06% URL + 0.1% Span) |
All other ranges | ±(0.04% URL + 0.065% Span) |