ABB 2600T/264 Pressure Transmitter
| Type | Data source |
|---|---|
| 264GS | Product Data Sheet SS/264GS/AS_4 2006 |
| 264PS | Product Data Sheet SS/264XS_6 2006 |
| 264PF | Product Data Sheet SS/264XF_5 2006 |
Reference Accuracy U(ac)
264GS
| Range | Low Turndown | High Turndown |
|---|---|---|
| C | For spans less than 10:1 ± 0.075% of calibrated span |
For spans from 10:1 to 30:1 |
| All other ranges | For spans less than 15:1 ± 0.075% of calibrated span |
For spans from 15:1 to 60:1
|
264PS ¶
| Range | Low Turndown | High Turndown |
|---|---|---|
| For spans less than 4:1 ± 0.1% of calibrated span |
For spans from 4:1 to 20:1
|
|
| B | For spans less than 10:1 ± 0.075% of calibrated span |
For spans from 10:1 to 20:1
|
| All other ranges | For spans less than 15:1 ± 0.075% of calibrated span |
For spans from 15:1 to 60:1
|
264PF ¶
| Range | Low Turndown | High Turndown |
|---|---|---|
| B | For spans less than 10:1 ± 0.075% of calibrated span |
For spans from 10:1 to 20:1
|
| All other ranges | For spans less than 15:1 ± 0.075% of calibrated span |
For spans from 15:1 to 60:1![]() |
Drift (stability) U(d)
264GS
± 0.15% of URL for 5 years
264PS
| Range A | ± 0.25% of URL for 5 years |
|---|---|
| All other ranges | ± 0.15% of URL for 5 years |
264PF
± 0.15% of URL for 5 years
Power supply effect U(ps)
All Models
± 0.005% of calibrated span per volt change.
Ambient temperature effect U(ac)
264GS
| Range | |
|---|---|
| C | For spans less than 10:1 ±(0.06% URL + 0.1% Span) |
| All other ranges | For spans less than 15:1 ±(0.04% URL + 0.065% Span) |
264PS
| Range | |
|---|---|
| A | For spans less than 4:1 ±(0.1% URL + 0.1% Span) |
| B | For spans less than 10:1 ±(0.06% URL + 0.1% Span) |
| All other ranges | For spans less than 15:1 ±(0.04% URL + 0.065% Span) |
264PF
| Range | |
|---|---|
| B | For spans less than 10:1 ±(0.06% URL + 0.1% Span) |
| All other ranges | For spans less than 15:1 ±(0.04% URL + 0.065% Span) |