Foxboro IGP20 Pressure Transmitter
| Type | Data source |
|---|---|
| IGP20 Analogue | Product Data Sheet PSS 2A-1C13 C |
| IGP20 HART | Product Data Sheet PSS 2A-1C13 B |
| IGP20 Fieldbus | Product Data Sheet PSS 2A-1C13 E |
Reference Accuracy U(ac)
Analogue
± 0.2% of calibrated span
HART
| Range | Spans < 10% URL | Spans > 10% URL |
|---|---|---|
| E & F | ![]() |
± 0.08% of calibrated span |
| All other ranges | ![]() |
± 0.06% of calibrated span |
Fieldbus
| Range | Spans > 10% URL |
|---|---|
| E & F | ± 0.07% of calibrated span |
| All other ranges | ± 0.05% of calibrated span |
Drift (stability) U(d)
All Models
± 0.05% of URL for 5 years
Power supply effect U(ps)
All Models
± 0.005% of calibrated span per volt change.
Ambient temperature effect U(ac)
Analogue
| Range | |
|---|---|
| A | ±(0.18% URL + 0.15% Span) |
| B & C | ±(0.03% URL + 0.2% Span) |
| D | ±(0.05% URL + 0.18% Span) |
| E & F | ±(0.08% URL + 0.15% Span) |
HART
| Range | |
|---|---|
| A | ±(0.18% URL + 0.025% Span) |
| B & C | ±(0.03% URL + 0.06% Span) |
| D | ±(0.05% URL + 0.045% Span) |
| E & F | ±(0.08% URL + 0.025% Span) |
Fieldbus
±(0.08% URL + 0.025% Span)